Erklärung von Blende - objektiv aufbau
Class3Rlaser
Foreseeable conditions are prepared for during normal operation. The limit value of accessible radiation in accordance with DIN EN 60825-1: 2001-11 in the wavelength range 400 nm to 1400 nm for the classification of lasers is the same between 100 s and 30,000 s.
Iris diaphragm: a unit below the condenser that controls the amount of light directed to the specimen. The diameter of the diaphragm can be adjusted by turning it to increase or decrease the size of the hole that light passes through.
Ocular lens or eyepiece: the secondary optical system that you look through. The ocular lens further magnifies (10x) the image and brings the light rays to a focal point. A binocular microscope has two ocular lenses and a monocular microscope has one ocular lens that sit on the adjustable binocular body. Binocular lenses can be adjusted to fit the distance between your eyes by gently pulling the oculars apart or by pushing them closer together.
The accessible laser radiation is only in the visible spectral range (400 nm to 700 nm). It is harmless to the eye for a short exposure time (up to 0.25 s). The eye is protected by the eyelid closure reflex when looking into the laser radiation for a short time. For continuous beam lasers of class 2, the limit value of accessible radiation (GZS) is 1 mW (for C6 equals 1).
The accessible laser radiation is in the visible spectral range of 400 nm to 700 nm. Short periods of exposure (up to 0.25s) are not dangerous to the eye, as long as the laser cross-section is not reduced by optical instruments (magnifying glasses, lenses, telescopes)! Additional radiation components outside the wavelength range of 400-700 nm fulfill the conditions for class 1M.
The accessible laser radiation is only in the visible spectral range (400 nm to 700 nm). It is harmless to the eye for a short exposure time (up to 0.25 s), which is protected by the eyelid closure reflex when looking into the laser radiation for a short time at random. For continuous beam lasers of class 2M, the limit value of accessible radiation (GZS) is 1 mW (for C6 equals 1).
When setting up and working within the working range of the laser beam, suitable protective clothing, laser safety glasses and gloves must be worn. The working range of the laser beam must be indicated by suitable warning signs. Laser equipment must be protected against unauthorized use. An additional visual warning lamp “Laser in operation” must be used to signal when the laser is switched on. The warning lamp must be clearly visible even from a greater distance. The laser beam must not scatter outside the intended working area. There must be no reflective surfaces in the working area.
Köhler illumination is the alignment of the image-forming light path and the illumination light path of the microscope. In this process the con-denser is centered and focused, thereby providing an evenly illuminated field of view and more importantly maximum resolution of the specimen
The accessible laser radiation is very dangerous for the eye and dangerous for the skin. Diffuse scattered radiation can also be dangerous. The laser radiation can cause fire and explosion hazards.
5 times the limit value of laser class 2 in the visible wavelength range 5 times the limit value of laser class 1 in the remaining wavelength range
Depth of Field: is determined by the distance from the nearest specimen plane in focus to that of the farthest plane also simultaneously in focus. The thickness of the optical section along the optical axis within which objects in the specimen plane are in focus. High-magnification objectives have a decreased depth of field. The reverse is true of low-magnification objectives Field of View: the visible area seen through the microscope when the specimen is in focus. The greater the magnification the smaller the view. Focus: a specimen is in focus at the desired magnification when the image seen through the ocular lens is sharp and clear.
Class2laserpointer
Z-LASER manufactures its products in accordance with international standards and declares the laser classes of its products in accordance with EN 60825-1, IEC 825-1 and 21 CFR 1040. The laser class can be found on the yellow / black / white / red (USA) warning sticker on each laser or on its firmly secured dimming device.
Base: the bottom of the microscope, which supports the entire instrument. The stage plate is located directly on the base surface upon which a specimen is placed. The stage can have a removable black or white tile (that can be removed and cleaned) or it will have a light that will transmit light through the specimen.
Coarse adjustment or coarse focusing knob: the large knob towards the back of the instrument that is used to significantly raise or lower the stage, when you first focus on a specimen at low power. It is never used when high power objectives are in place.
Class3Blaser
Illuminator or light source: the light source can be built into the base of the microscope, transmitting light through the specimen and/or the light source may be above the specimen as incident light. The lights can be turned on using rheostat (light) control knob on the front of the base.
Laserclasses
Class 2M laser devices may be used without further protective measures if it is ensured that it is neither possible to look into the laser radiation intentionally for longer than 0.25 s nor to look repeatedly into the laser radiation or directly reflected laser radiation. In addition, it must be ensured that no optical collecting instruments are used in the area of projection. In the case of Class 2M laser equipment, there is generally no laser area requiring additional protection if only random irradiation of persons is possible when operating this laser equipment and no optically collecting instruments are used.
*) Note on laser classes 2 and 2M: Scientific investigations (Cologne University of Applied Sciences) found that the eyelid closure reflex (which fortunately occurs within 0.25 s, as prolonged radiation damages the eye) was only present in <20% of the test subjects. The presence of the eyelid reflex for the protection of the eyes can therefore generally not be relied upon. Therefore, if class 2 or 2M laser radiation hits the eye, one should close the eyes or turn away immediately. Furthermore, it should be noted that the eyelid closure reflex occurs only in cases of visible light. Laser radiation in the infrared range, for example, does not trigger eyelid closure, as it is not perceived by the eye. Therefore, particularly careful handling of invisible laser radiation is recommended.
For class 3R continuous beam lasers, the limit of accessible radiation (GZS) is 5 mW (for C6 equal to 1) in the visible wavelength range.
Microscope are used by the students in many lab exercises. Instructors also need to learn to use the instructor microscope with the Leica camera and required LAS EZ & Leica AirLab Icon Guide software which will allow them to project the microscope images in real time.
Stage: the flat surface upon which the slide with your specimen is placed. Most microscopes have a stage finger assembly to hold the slide on the stage. The entire mechanism including the slide moves horizontally across the stationary stage (left/right and forward/back) using two stage adjustment knobs situated under the stage (variably on the left or right side, in front of the focusing knobs).
Class 2 laser devices may be used without further protective measures, provided that it is ensured that it is neither possible to look into the laser radiation intentionally for longer than 0.25 s nor to look repeatedly into the laser radiation or directly reflected laser radiation. In the case of Class 2 laser equipment, there is generally no laser area requiring additional safety measures if the operation of such laser equipment allows only accidental irradiation of persons.
Illuminator or light source: the light source is usually built into the base of the microscope, and directs light through the condenser to the specimen.Alternatively, the light source may be separate, and be directed toward the condenser with a mirror. The intensity of the light can be adjusted using the rheostat (light) control knob. The microscope you are using has a rheostat on the front of the base and a switch on the left of the base.
The 100X objective lens is called an oil immersion lens because oil is placed between the lens and the microscope slide to increase resolution (i.e., the level of detail that can be observed in an image). Light bends when it passes from the glass slide to air because of differing refractive indices. A drop of immersion oil between the slide and lens eliminates this problem because the oil has the same refractive index as the glass slide. Never use the 100X objective lens without oil and do not get oil on the 4X, 10X, or 40X lenses.
Class1laser
Unless optical instruments are used which reduce the beam cross-section, there is a comparable hazard for class 2M laser equipment as for class 2 laser equipment.
Class1Mlaser
Axial Resolution: point-to-point resolving power in the plane parallel to the optical axis. It is usually defined at the shortest distance between two longitudinal points on the specimen plane that can still be distinguished as separate entities.
Microscopes must be calibrated so accurate measurements can be made. To calibrate a microscope both an ocular and a stage micrometer are used.
Focusing knob: the knob that allows you to focus on the object at each magnification by moving the stereo head up or down.
Note: The microscope is now set to maximize resolution of the specimen. If you adjust the condenser height to gain contrast or adjust light intensity you will sacrifice the resolution capability. Use the aperture diaphragm and /or the illumination intensity to adjust contrast.
Laser class 3R devices are potentially dangerous to the eyes. The risk of eye damage is reduced by the limit value of the maximum accessible radiation (GZS) of:
Condenser: the lens located below the stage, which focuses light (from the illuminator) through the specimen being observed. Most microscopes have a movable condenser allowing its distance from the specimen to be adjusted using the condenser knob and condenser alignment screws.
A compound microscope is a high power microscope that uses a compound lens system. Higher magnification is achieved by using two lenses rather than just a single magnifying lens. While the eyepieces and the objective lenses create high magnification, a condenser beneath the stage focuses the light directly into the sample. A compound microscope has multiple lenses: the objective lens (typically 4x, 10x, 40x or 100x) is compounded (multiplied) by the eyepiece lens (typically 10x) to obtain a high magnification of 40x, 100x, 400x and 1000x. The objective lenses of a compound microscope causes the orientation of the image of the specimen to be inverted compared to the orientation of the actual specimen which means that a specimen viewed through a compound microscope will look upside down and backwards compared to how the specimen is mounted on the slide.
Laser class 3B is dangerous for the eyes and potentially also for the skin. Persons must not be exposed to the laser beam. Diffusely scattered laser radiation is not dangerous as long as the viewing distance of 13 cm and a viewing time of less than 10 s is guaranteed. There is a fire hazard if the laser beam comes into contact with flammable materials.
Lateral Resolution: point-to-point resolving power in the plane perpendicular to the optical axis. It is usually defined as the shortest distance between two lateral points on the specimen plane that can still be distinguished as separate entities.
Class4laser
Objective lenses: the primary optical system which produces a magnified image of the specimen. There are typically four objective lenses attached to the nosepiece with the magnification of each objective is engraved on its side.
The accessible laser radiation is very dangerous for the eye and dangerous for the skin. Diffuse scattered radiation can also be dangerous. The laser radiation can cause fire and explosion hazards. Note: Class 4 laser devices are high-power lasers whose output powers or energies exceed the limits for accessible radiation (GZS) for Class 3B. The laser radiation from Class 4 laser devices is so intense that damage to the eyes or skin is to be expected from any type of exposure. If necessary, secure access to the room by sluices or door contacts (laser shut down) if the degree of danger so requires. In addition, when using class 4 laser equipment, it must always be checked whether sufficient measures against fire and explosion hazards have been taken; see also § 10 Laser radiation (previously: BGV B2) and § 16 Laser radiation (previously: BGV B2) of the accident prevention regulation Laser radiation.
Stereo microscopes have low magnifications that can range from 2 to 100x depending on the microscope, and are designed for viewing whole objects like rocks, plants, flowers, and invertebrate organisms by reflecting light off the specimen, producing a 3-dimensional image. Sometimes there is a light located in the base of the microscope that will allow transmitted light.
The resolving power of a microscope is dependent on the numerical apertures of the optical lenses and the wavelength of light used to examine the specimen. It is the smallest distance between two points (measured in microns) that can be seen with the microscope. If two small objects close together can be seen clearly as two distinct objects, a microscope is said to have high resolving power.
Class 2m laserpointer
Diopter: compensates for focusing differences between your eyes, it is very important this is set correctly, in order to prevent eye strain.
Accidental reflections must be avoided. Beam indicators, interlock and key switch are not required in the visible wavelength range. No special requirements regarding eye protection devices and protective clothing are necessary. Operating and maintenance personnel must be trained in handling laser equipment.
Fine adjustment or fine focusing knob: the smaller knob towards the back of the instrument that is used to make small adjustments in the height of the stage for final focusing on a specimen. It is the only focusing knob used with high power objectives.
Magnification is the process of enlarging the apparent size, not physical size, of something. In microscopy, it is the ratio between the size of an image produced by the microscope and its actual size. Microscopes magnify thin specimens mounted on microscope slides. They are ideal for observing unicellular or very small organisms, cells, and cell structures. We will use the compound and dissecting microscopes many times over the course of the semester. It is important to familiarize yourself with microscope use.